Resolution (electron density)

Resolution in terms of electron density is a measure of the resolvability in the electron density map of a molecule. In X-ray crystallography, resolution is the highest resolvable peak in the diffraction pattern. While cryo-electron microscopy is a frequency space comparison of two halves of the data, which strives to correlate with the X-ray definition.[1]

Contents

History

Qualitative measures

In structural biology, resolution can be broken down into 4 groups:

X-ray crystallography

As the crystal's repeating unit, its unit cell, becomes larger and more complex, the atomic-level picture provided by X-ray crystallography becomes less well-resolved (more "fuzzy") for a given number of observed reflections. Two limiting cases of X-ray crystallography are often discerned, "small-molecule" and "macromolecular" crystallography. Small-molecule crystallography typically involves crystals with fewer than 100 atoms in their asymmetric unit; such crystal structures are usually so well resolved that its atoms can be discerned as isolated "blobs" of electron density. By contrast, macromolecular crystallography often involves tens of thousands of atoms in the unit cell. Such crystal structures are generally less well-resolved (more "smeared out"); the atoms and chemical bonds appear as tubes of electron density, rather than as isolated atoms. In general, small molecules are also easier to crystallize than macromolecules; however, X-ray crystallography has proven possible even for viruses with hundreds of thousands of atoms.  

A rough guide to the resolution of protein structures
Resolution (Å) Meaning
>4.0 Individual coordinates meaningless
3.0 - 4.0 Fold possibly correct, but errors are very likely. Many sidechains placed with wrong rotamer.
2.5 - 3.0 Fold likely correct except that some surface loops might be mismodelled. Several long, thin sidechains (lys, glu, gln, etc.) and small sidechains (ser, val, thr, etc.) likely to have wrong rotamers.
2.0 - 2.5 As 2.5 - 3.0, but number of sidechains in wrong rotamer is considerably less. Many small errors can normally be detected. Fold normally correct and number of errors in surface loops is small. Water molecules and small ligands become visible.
1.5 - 2.0 Few residues have wrong rotamer. Many small errors can normally be detected. Folds are extremely rarely incorrect, even in surface loops.
0.5 - 1.5 In general, structures have almost no errors at this resolution. Rotamer libraries and geometry studies are made from these structures.

Cryo-electron microscopy

In cryo-electron microscopy, resolution is typically measured by the Fourier shell correlation (FSC),[2] a three-dimensional extension of the Fourier ring correlation (FRC).[3][4] The FSC is a comparison of two different Fourier transforms over different shells on frequency space. To measure the FSC, the data needs to be separated into two groups. Typically, the even particles form the first group and odd particles the second based on their order. This is commonly referred to as the even-odd test. Most publications quote the FSC 0.5 cutoff, which refers to when the correlation coefficient of the Fourier shells is equal to 0.5.[1][5]

Determining the resolution threshold remains a controversial topic and many other criteria using the FSC curve exist, including 3-σ criterion, 5-σ criterion, and the 0.143 cutoff. However, fixed-value thresholds (like 0.5, or 0.143) were argued to be based on incorrect statistical assumptions.[6] The new half-bit criterion indicates at which resolution one has collected enough information to reliably interpret the 3-dimensional volume, and the (modified) 3-sigma criterion indicates where the FSC systematically emerges above the expected random correlations of the background noise.[6] In 2007, a resolution criterion independent of the FSC was developed using the correlation between neighboring Fourier to distinguish signal from noise.[7]

Notes

  1. ^ a b Frank, 2006, p250-251
  2. ^ Harauz & van Heel, 1986
  3. ^ van Heel, 1982
  4. ^ Saxton & Baumeister, 1982
  5. ^ Böttcher et al., 1997
  6. ^ a b van Heel & Schatz, 2005
  7. ^ Sousa & Grigoreiff, 2007

References

  • Harauz, G.; M. van Heel (1986). "Exact filters for general geometry three dimensional reconstruction". Optik 73: 146–156. 
  • van Heel, M. (1982). "Detection of objects in quantum-noise limited images". Ultramicroscopy 8: 331–342. 
  • Saxton, W.O.; W. Baumeister (1982). "The correlation averaging of a regularly arranged bacterial cell envelope protein". J Microscopy 127: 127–138. 
  • Böttcher, B.; Wynne, S.A.; Crowther, R.A. (1997). "Determination of the fold of the core protein of hepatitis B virus by electron microscopy". Nature 386 (6620): 88–91. doi:10.1038/386088a0. PMID 9052786. 
  • van Heel, M.; Schatz, M. (2005). "Fourier shell correlation threshold criteria". Journal of Structural Biology 151 (3): 250–262. doi:10.1016/j.jsb.2005.05.009. PMID 16125414. 
  • Frank, Joachim (2006). Three-Dimnsional Electron Microscopy of Macromolecular Assemblies. New York: Oxford University Press. ISBN 0-19-518218-9.